Sacituzumab govitecan (anti-TROP2) ELISA Kit

Referencia EU6794-96T

embalaje : 96T

Marca : FineTest

Contact local distributor :


Teléfono : +1 850 650 7790

This kit is based on indirect ELISA detection method and takes 3h assay time. The microplate provided in this kit has been precoated with antigen. HRP-antibody is used as detection antibody. Add standard and properly diluted sample into relevant well respectively. After incubation, wash unbound components. Add HRP-detection antibody, then it binds with Sacituzumab govitecan bound to precoated antibody. Wash unbound components and add TMB substrate solution. Then, TMB was catalyzed by HRP to produce a blue color product that turned yellow after adding a stop solution. Read the O.D. absorbance at 450nm in a microplate reader. Calculate the concentration of Sacituzumab govitecan in the sample by plotting standard curve. The concentration of the target substance is proportional to the OD450 value.

Catalogue No.: 
EU6794
Alias: 
Sacituzumab govitecan ELISA Kit, IMMU-132 ELISA Kit
Species: 
Universal
Range: 
0.781-50ng/ml
Sensitivity: 
0.469ng/ml
Inquiry
Dispatch Time: About 3 working days
Product Name
Sacituzumab govitecan (anti-TROP2) ELISA Kit
Alias
Sacituzumab govitecan ELISA Kit, IMMU-132 ELISA Kit
Catalogue No.
EU6794
Size
96T
Species
Universal
UniProt ID
CAS No. : 1491917-83-9
Sample Type
Serum, Plasma, Cell Culture Supernatant, cell or tissue lysate, Other liquid samples
Detection Method
Indirect ELISA
Detection Wavelength
OD450
Reaction Duration
3 hours
Range
0.781-50ng/ml
Sensitivity
0.469ng/ml
Storage
2-8°C(Sealed), Don't cryopreserve.
Specificity
Specifically binds with Sacituzumab govitecan , no obvious cross reaction with other analogues.
Required Instruments and Reagents
  1. Microplate reader (wavelength: 450nm)
  2. 37°C incubator (CO2 incubator for cell culture is not recommenced.)
  3. Automated plate washer or multi-channel pipette/5ml pipettor (for manual washing purpose)
  4. Precision single (0.5-10μL, 5-50μL, 20-200μL, 200-1000μL) and multi-channel pipette with disposable tips(Calibration is required before use.)
  5. Sterile tubes and Eppendorf tubes with disposable tips
  6. Absorbent paper and loading slot
  7. Deionized or distilled water
Assay Procedure Summary
Elisa实验原理图
  • Step 1: Wash the plate twice before adding the standard and sample.
  • Step 2: Add 100ul standard or sample into each well, seal the plate and statically incubate for 90 minutes at 37°C.
  • Washing: Wash the plate three times and immerse for 1min each time.
  • Step 3: Add 100ul HRP-antibody working solution, seal the plate and statically incubate for 30 minutes at 37°C.
  • Washing: Wash the plate five times and immerse for 1min each time.
  • Step 4: Add 90ul TMB substrate solution, seal the plate and statically incubate for 10-20 minutes at 37°C. (Accurate TMB visualization control is required.)
  • Step 5: Add 50ul stop solution. Read at 450nm immediately and calculate.
Standard Curve

This product is detected by QC department and meets performance required in the manual. (Laboratory Humidity: 20%-60%; Temperature: 18°C -25°C; Equilibrate TMB substrate to 37°C before staining. After adding into the ELISA wells, incubate for 15min at 37°C in dark.)

Due to different assay environments and operations, assay data below and standard curve are provided for reference. Experimenters should establish standard curve according to their own assay.

STD.(ng/ml) OD-1 OD-2 Average
0 0.084 0.086 0.085
0.781 0.129 0.133 0.131
1.563 0.246 0.254 0.25
3.125 0.396 0.408 0.402
6.25 0.548 0.564 0.556
12.5 0.775 0.797 0.786
25 1.387 1.427 1.407
50 2.049 2.109 2.079
EU6794 Standard Curve Image
Recovery

Add a certain amount of Sacituzumab govitecan into the sample. Calculate the recovery by comparing the measured value with the expected amount of Sacituzumab govitecan in the sample.

Sample Type Recovery Range(%) Average(%)
serum(n=10) 93-102 97
EDTA plasma(n=10) 93-103 98
Heparin plasma(n=10) 87-100 92
Linearity

Dilute the sample with a certain amount of Sacituzumab govitecan at 1:2, 1:4 and 1:8 to get the recovery range.

Sample Type 1:2 1:4 1:8
serum(n=10) 86-105% 85-97% 92-103%
EDTA plasma(n=10) 82-101% 91-98% 84-101%
Heparin plasma(n=10) 91-97% 86-96% 80-97%
Precision(%)

Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on the same plate.

Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.

Item Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (ng/ml) 1.64 6.34 25.2 1.58 6.55 24.64
Standard deviation 0.08 0.35 1.11 0.08 0.31 1.31
CV(%) 5.03 5.52 4.42 5.12 4.69 5.3
Stability

Perform the stability test for the sealed kit at 37°C and 2-8°C and get relevant data.

ELISA kit(n=5) 37°C for 1 month 2-8°C for 6 months 2-8°C for 12 months
Average(%) 80 95-100 85-98

Product Comparison

Usted podría estar interesado también en los siguientes productos:



Referencia
Descripción
Cond.
Precio Sin IVA