Rat Adiponectin ELISA Kit PicoKine®
Referencia EK1327
embalaje : 96wells/kit,withremovablestrips.
Marca : Boster Bio
Product Info Summary
| SKU: | EK1327 |
|---|---|
| Size: | 96 wells/kit, with removable strips. |
| Reactive Species: | Rat |
| Application: | ELISA |
| Sample Types: | cell culture supernatants, serum, plasma (heparin, EDTA) and urine. |
Product Name
Rat Adiponectin ELISA Kit PicoKine®
SKU/Catalog Number
EK1327
Size
96 wells/kit, with removable strips.
*Question: How many samples can I assay/run in this kit?
Description
Rat Adiponectin ELISA Kit PicoKine® (96 Tests). Quantitate Rat Adipoq in cell culture supernatants, serum, plasma (heparin, EDTA) and urine. Sensitivity: 10pg/ml. The brand Picokine indicates this is a premium quality ELISA kit. Each Picokine kit delivers precise quantification, high sensitivity, and excellent reproducibility. Only our most reliable and effective kits qualify as Picokine, guaranteeing top-tier results for your assays.
Storage & Handling
Store at 4°C for 6 months, at -20°C for 12 months. Avoid multiple freeze-thaw cycles (Ships with gel ice, can store for up to 3 days in room temperature. Freeze upon receiving.)
Cite This Product
Rat Adiponectin ELISA Kit PicoKine® (Boster Biological Technology, Pleasanton CA, USA, Catalog # EK1327)
Clonality of Antibodies
See Datasheet for details
Standard Protein
Expression system for standard: E.coli; Immunogen sequence: I19-N244
Sensitivity
<10 pg/ml
Assay Range
156 pg/ml - 10,000 pg/ml
Standard Dilution Instructions

See datasheet of EK1327 for more details
Cross-reactivity
There is no detectable cross-reactivity with other relevant proteins.
Reactive Species
EK1327 is reactive to Adipoq in Rat samples
Validated Sample Types
cell culture supernatants, serum, plasma (heparin, EDTA) and urine.
Application Guarantee
EK1327 is guaranteed for ELISA in Rat by Boster Guarantee
See how Boster Bio validate our ELISA kits: ELISA Validation Information
Background of Adipoq
Adiponectin (ADPN) is a hormone secreted by adipocytes that regulates energy homeostasis and glucose and lipid metabolism. Adiponectin is a new member of the family of soluble defense collagens, in hematopoiesis and immune responses. It is an important negative regulator in hematopoiesis and immune systems and raise the possibility that it may be involved in ending inflammatory responses through its inhibitory functions. Adiponectin is mapped to 3q27 and can protect the organism from systemic inflammation by promoting the clearance of early apoptotic cells by macrophages through a receptor-dependent pathway involving calreticulin.
Kit Components
| Catalog Number | Description | Quantity |
|---|---|---|
| EK1327-CAP | Anti-Rat Adipoq Pre-coated 96-well strip microplate | 1 |
| EK1327-ST | Rat Adipoq Standard | 2 vials, 10 ng/tube |
| EK1327-DA | Rat Adipoq Biotinylated antibody (100x) | 100ul |
| AR1103 | Avidin-Biotin-Peroxidase Complex (100x) | 100ul |
| AR1106-1 | Sample Diluent | 30ml |
| AR1106-2 | Antibody Diluent | 12ml |
| AR1106-3 | Avidin-Biotin-Peroxidase Diluent | 12ml |
| AR1104 | Color Developing Reagent (TMB) | 10ml |
| AR1105 | Stop Solution | 10ml |
| AR1106-5 | Wash Buffer (25x) | 20ml |
| PLA-SEA | Adhesive plate sealers | 4 |
*The kit components are not available for individual purchase.
Materials Required But Not Included In Kit
- Microplate Reader capable of reading absorbance at 450nm.
- Incubator.
- Automated plate washer (optional).
- Pipettes and pipette tips capable of precisely dispensing 0.5 µl through 1 ml volumes of aqueous solutions.
- Multichannel pipettes are recommended for large amount of samples.
- Deionized or distilled water.
- 500ml graduated cylinders.
- Test tubes for dilution.
Validation Standard Curve O.D. At 450nm
| Concentration (pg/ml) | 0 | 156 | 312 | 625 | 1250 | 2500 | 5000 | 10000 |
| O.D. | 0.017 | 0.102 | 0.149 | 0.224 | 0.471 | 0.837 | 1.416 | 1.923 |
Data Example Images
Recommended Sample Dilution Ratios
According to our internal validation assays using this ELISA kit, to detect Adipoq,
| Application | Dilution | Species |
|---|---|---|
| Dilution ratio of 1:2000 | concentration in serum and plasma is 2.41-8.136 ug/ml. |
Intra Assay Consistency & Inter Assay Consistency
We measured random samples of Rat Adiponectin ELISA Kit PicoKine® within the same batch/lot to ensure the consistency of the kits' performances. ELISA intra assay consistency is measured using wells from the same plate/assay kit. ELISA inter assay consistency is measured using wells from different plates from the same batch production/lot.
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 16 | 16 | 16 | 24 | 24 | 24 |
| Mean (pg/ml) | 292 | 1705 | 4269 | 265 | 1626 | 4075 |
| Standard deviation | 16.93 | 126.17 | 170.76 | 19.61 | 133.33 | 167.07 |
| CV (%) | 5.8% | 7.4% | 4% | 7.4% | 8.2% | 4.1% |