Human SARS-CoV-2 Nucleocapsid ELISA Kit
Referencia NB-66-32276-96T
embalaje : 96T
Marca : Neo Biotech
Products
Overview
Standard curve
Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant SARS-CoV-2 Nucleocapsid and the recovery rates were calculated by comparing the measured value to the expected amount of SARS-CoV-2 Nucleocapsid in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-105% | 97% |
| EDTA plasma(n=5) | 79-97% | 88% |
| Heparin plasma(n=5) | 80-92% | 86% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of SARS-CoV-2 Nucleocapsid and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-105% | 88-102% | 95-102% | 87-98% |
| EDTA plasma(n=5) | 79-96% | 85-99% | 98-105% | 92-101% |
| Heparin plasma(n=5) | 85-105% | 92-104% | 95-102% | 86-97% |