Guinea pig IL-12p40(Interleukin 12 p40) ELISA Kit
Cat# NB-66-26905-48T
Size : 48T
Brand : Neo Biotech
Products
Overview
Standard curve
Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant IL-12p40 and the recovery rates were calculated by comparing the measured value to the expected amount of IL-12p40 in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 79-93% | 86% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 85-95% | 90% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of IL-12p40 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-104% | 83-96% | 95-103% | 88-95% |
| EDTA plasma(n=5) | 88-97% | 91-103% | 86-92% | 89-97% |
| Heparin plasma(n=5) | 85-94% | 87-96% | 89-101% | 82-90% |