Goat Luteinizing Hormone (LH) ELISA KIT
Referência CSB-E13274G-5x96T
Tamanho : 5x96T
Marca : Cusabio
| Size | 96T,5×96T,10×96T |
Product Details
Abbreviation
LH
Species
Capra hircus (Goat)
Sample Types
serum, plasma, tissue homogenates
Detection Range
0.3 mLU/mL-60 mLU/mL
Sensitivity
0.15 mIU/mL
Assay Time
1-5h
Sample Loading Volume
50-100ul
Detection Wavelength
450 nm
Research Area
Signal Transduction
Assay Principle
quantitative
Measurement
Competitive
Precision
| Intra-assay Precision (Precision within an assay): CV%<15% | ||||||
| Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
| Inter-assay Precision (Precision between assays): CV%<15% | ||||||
| Three samples of known concentration were tested in twenty assays to assess. | ||||||
Linearity
| To assess the linearity of the assay, samples were spiked with high concentrations of goat LH in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
| Sample | Serum(n=4) | |||||
| 1:1 | Average % | 98 | ||||
| Range % | 90-104 | |||||
| 1:2 | Average % | 95 | ||||
| Range % | 90-103 | |||||
| 1:4 | Average % | 91 | ||||
| Range % | 85-96 | |||||
| 1:8 | Average % | 90 | ||||
| Range % | 88-93 | |||||
Recovery
| The recovery of goat LH spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
| Sample Type | Average % Recovery | Range | ||||
| Serum (n=5) | 96 | 90-101 | ||||
| EDTA plasma (n=4) | 92 | 87-96 | ||||
Typical Data
| These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||
ELISA Data Analysis
Troubleshooting
and FAQs
and FAQs
Storage
Store at 2-8°C. Please refer to protocol.
Shelf Life
6 months
Description
This Goat Luteinizing Hormone (LH) ELISA Kit was designed for the quantitative measurement of Goat Luteinizing Hormone (LH) protein in serum, plasma, tissue homogenates. It is a Competitive ELISA kit, its detection range is 0.3 mLU/mL-60 mLU/mL and the sensitivity is 0.15 mIU/mL .
Citations
- Y Ding,/,2023
- Arjmand M et al,Vet Res Forum,2014
Customer Reviews and Q&A
There are currently no reviews for this product.